-40%

Sencorp White Ceratek 24" Variable Medical Pouch Sealer 24-PV/2

$ 8395.19

Availability: 100 in stock
  • Condition: Late model units with LOW usage in excellent condition. Manufacture dates are in 2019/ 2018. Cycle counts on the machines are very low at: 48, 65, and 338. Each unit comes with a power cord. 3 units are available, each bid is for ONE (1) unit. Foot pedals are NOT included, however I do have one available, please contact me for pricing.
  • Return shipping will be paid by: Buyer
  • Item must be returned within: 30 Days
  • All returns accepted: Returns Accepted
  • MPN: 24-PV/2
  • Model: Ceratek 24-PV/2 24" Medical Pouch Sealer
  • Refund will be given as: Money Back
  • Brand: Sencorp White

    Description

    Sencorp White
    Ceratek 24" Variable Medical Pouch Sealer
    Model:
    24-PV/2
    Condition:
    Late model units
    with
    LOW usage
    in
    excellent condition
    .  Manufacture dates are in 2019/ 2018.
    Cycle counts
    on the machines are very low at:
    48, 65, and 338
    .  Each unit comes with a power cord. 3 units are available, each bid is for ONE (1) unit.
    Foot pedals are NOT included, however I do have one available, please contact me for pricing.
    24" wide seal jaw
    Single Phase 120VAC
    ISO 11697
    CE compliant
    Configurable risk mitigation
    Barcode scanning options
    Info from Manufacturer
    :
    A
    bar sealer
    with
    top heat
    ,
    process control
    and display through the HMI,
    surge tank
    ,
    validation ports
    ,
    process fault indication
    and
    vacuum/gas flush capabilities.
    The P-series control system is PLC based with an LCD touchscreen HMI. All process values and set-points are displayed and modified through the HMI. Process parameters and fault reset controls are accessible with a key switch to prevent unauthorized modifications.
    Seal time is controlled by the PLC. The range is 0.10 – 99.99 seconds.
    If the cycle is aborted before the dwell time elapses a “short cycle” message is displayed and the machine is disabled until the fault is cleared.
    Closed loop temperature control utilizes thermocouple feedback to a PLC module. A temperature deviation process alarm is monitored by the PLC. Both high and low temperature limits can be entered through the HMI.
    If the temperature falls outside the acceptable process window a message is displayed and the machine is disabled until the fault is cleared. Jaw pressure is controlled manually with a precision regulator. A pressure transducer sends a signal to the PLC to monitor the actual pressure. Both high and low- pressure alarms can be entered through the HMI. If the pressure falls outside the acceptable process window a message is displayed and the machine is disabled until the fault is cleared.
    In
    “PV” models,
    an onboard vacuum source is used to evacuate a pouch. The operation mode of the vacuum system can be set to either “time” or “level”. If the operation mode is set to “time” the vacuum turns on for a set period of time after the operator initiates the cycle. The operator can adjust the time as required. If the vacuum mode is set to “level” the vacuum stays on until the vacuum level equals the process set point. The vacuum level is monitored in the PLC with a signal from a pressure transducer. When either the “dwell time” ends or the required level is reached the vacuum valve turns off and the nozzle retracts. If the required vacuum level is not achieved in the “level” mode a message is displayed and the machine is disabled until the fault is cleared.
    Worldwide Shipping Available
    NOTE:
    2 units can ship on a single pallet to save $$$ on shipping
    Shipping weight: 250LB per unit
    Shipping dimensions: 48" x 40" x 30"